<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>40516</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: Load table unable to measure stack shift.  Plates in cassette &lt;Cassette Number&gt; larger than expected.\nVerify correct size, orientation, and stack location.
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Magnus_VLF</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The loadtable was not able to measure the stackshift due to the plate already covering the sensor at the measure position.
This is likely due to an incorrect plate size, or an incorrect plate orientation.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Correct the plate size/orientation.
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Plate stack not centered
</Cause>
<Cause>
Incorrect media size.
</Cause>
<Cause>
Incorrect media orientation.
</Cause>
<Cause>
Faulty sensor.
</Cause>
<Parameter>
Cassette Number
</Parameter>
<LastModified>
26-Aug-2004
Created.
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:28 2008 PST
</Updated>
<Links>
</Links>
</Error>